| | SVM 2001 | SVM 3001 | SVM 3001 Cold Properties | SVM 4001 |
| Patents | AT516058 (B1), US10036695 (B2), CN105424556, EP2995928 (B1) | AT516058 (B1), US10036695 (B2), CN105424556, EP2995928 (B1) AT516302 (B1), CN105628550 |
| Test methods | ASTM D7042, D445 bias-corrected, ISO 23581, ASTM D4052 (optional) | ASTM D7042, D445 bias-corrected, ISO 23581, ASTM D4052, ISO 12185 | ASTM D7042, D445 bias-corrected, ISO 23581, ASTM D4052, ISO 12185 ASTM D2386 equiv. or better ASTM D2500 equiv. or better | ASTM D7042, D445 bias-corrected, ISO 23581, ASTM D4052, ISO 12185 |
| Measuring Range |
| Viscosity | 0.2 mm²/s to 30,000 mm²/s* |
| Density | 0 g/cm³ to 3 g/cm³ |
| Temperature | +15 °C to +100 °C | -60 °C to +135 °C | -60 °C to +100 °C | +15 °C to +100°C |
| Precision |
| Viscosity repeatability** and reproducibility** | 0.1 % and 0.35 % |
Density repeatability** and reproducibility** | 0.0002 g/cm³ (0.00005 g/cm² with ASTM D4052 option) and 0.0005 g/cm³ (0.0001 g/cm² with ASTM D4052 option) | 0.00005 g/cm³ and 0.0001 g/cm³ |
Cloud/freeze point repeatability** and reproducibility** | - | - | <0.5 °C / <0.5 °C and <2.5 °C / <1.3 °C | - |
| Temperature repeatability | 0.005 °C |
| Temperature reproducibility | 0.03 °C from +15 °C to +100 °C | 0.03 °C from +15 °C to +100 °C | 0.03 °C from +15 °C to +100 °C |
| Performance |
| Sample volume min. / typical | 1.5 mL / 5 mL | 2.5 mL / 6 mL |
| Solvent volume min. / typical | 1.5 mL / 6 mL | 2.5 mL / 10 mL |
| Max. sample throughput | 33 samples per hour | 30 samples per hour | 24 samples per hour |
| Features |
| Optional upgrades | Automatic VI determination ASTM D4052-compliant density measurement | Chemical upgrade kit | - | - |
| Modularity with Abbemat refractometers 5001, 5101, 5201, 7001, and 7201 |
| Accessories | Magnetic particle trap (MPT), hot filling attachment (HFA) | Magnetic particle trap (MPT) |
| Automation | Optional sample changer + High-Throughput Platform HTV | Optional sample changer + High-Throughput Platform HTV | Optional sample changer + High-Throughput Platform HTV |
| Technical Data |
| Data memory | 1,000 measurement results |
| HID (Human Interface Device) | 10.4" touchscreen, optional keyboard, mouse, and 2D bar code reader |
| Interfaces | 4 x USB (2.0 full speed), 1 x Ethernet (100 Mbit), 1 x CAN bus, 1 x RS-232, 1 x VGA |
| Power supply | AC 100 V to 240 V, 50 Hz to 60 Hz, 250 VA max. |
| Ambient conditions | 15 °C to 35 °C, max. 80 % r.h. non-condensing |
| Net weight in kg | 15.9 kg | 17.6 kg | 18.0 kg | 17.8 kg |
Dimensions (W x D x H) | 33 cm x 51 cm x 23.1 cm |
| Trademarks | SVM (13411996), FillingCheck (6834725), Abbemat (1084545) |