此为必填字段。
Invalid
零件型号
验证错误!

包装总氧测量仪:
TPO

  • Freestanding Anton Paar TPO 3001 Total Package Oxygen Meter with visible touchscreen and enclosed sample chamber for beverage oxygen analysis.
  • Close-up of the automatic piercing device mounted on the Anton Paar TPO 5001 for oxygen measurement in sealed beverage containers.
  • Technician loading a beverage can into the Anton Paar TPO 5001 Total Package Oxygen Meter; various bottles and cans shown for compatibility.
  • Close-up of the status indicator light on the Anton Paar TPO 5001, illuminated green to signal operational readiness.
  • Gloved hand adjusting temperature correction settings on the touchscreen of the Anton Paar TPO 5001 Total Package Oxygen Meter.
  • Lab technician loading a beer bottle into the Anton Paar TPO 5001; DMA 4501 M and PFD filling device visible in the background.
  • Freestanding Anton Paar TPO 3001 Total Package Oxygen Meter with visible touchscreen and enclosed sample chamber for beverage oxygen analysis.
  • Close-up of the automatic piercing device mounted on the Anton Paar TPO 5001 for oxygen measurement in sealed beverage containers.
  • Technician loading a beverage can into the Anton Paar TPO 5001 Total Package Oxygen Meter; various bottles and cans shown for compatibility.
  • +3
  • 360°
型号:
  • 可选择性测量易拉罐、玻璃瓶中的氧气总含量
  • 只要四分钟即可获得结果
  • 自清洁功能,只需极少的维护
  • 模块化扩展能满足多达 50 个行业特定测量参数
  • 在源头检测 O2 渗入,确保灌装机性能稳定
  • 一键测量,自动清洁
  • 模块化集成,可分析多达 50 个参数
  • 可选 360° 状态指示灯,用于监控测量进度
  • 选择性 CO2 分析升级,可满足扩展需求 

TPO 3001 和 5001 可选择性测量易拉罐、玻璃瓶或 PET 瓶中的顶空氧和溶解氧,精确测定饮料中的总氧含量。此类先进的包装总氧 (TPO) 测量仪专为在灌装线上使用而设计,可确保灌装机性能稳定,并保护产品的保质期、风味和外观。搭配安东帕久经验证的选择性 CO₂ 测量技术,模块化 TPO 测量仪系列可快速输出精准结果,并无缝集成至行业领先的多参数检测系统。

关键功能

精确定位氧气渗入位置

TPO 测量仪系列将实验室级精度带到生产线现场,数分钟内即可出具精准、可重复的 TPO 结果(含顶空氧与溶解氧)。通过定位氧气渗入位置,助力维持灌装机稳定运行,减少生产浪费。坚固的 IPX3 防护设计,确保在严苛环境下仍可稳定运行。为实现超高精度,TPO 测量仪系列可搭载选配的选择性 CO2 测量模块,其原理基于安东帕 20 余年持续优化的成熟技术。

Technician loading a beverage can into the Anton Paar TPO 5001 Total Package Oxygen Meter; various bottles and cans shown for compatibility.

快速、自动化、智能化的自我监测

TPO 测量仪系列的设计旨在提高效率和易用性,集快速检测、自动化运行与智能控制于一体,确保灌装机现场检测结果稳定可靠。配备一键测量功能,最快四分钟即可出具 TPO 检测结果。每次检测后传感器自动清洁,并持续自监测关键运行参数(如包装定位、气路与水路状态)—— 确保运行可靠无误。出现偏差时,操作员会立即收到屏幕指引,可选的 360° 状态指示灯可从任何角度跟踪测量结果和仪器状况。 

Close-up of the status indicator light on the Anton Paar TPO 5001, illuminated green to signal operational readiness.

轻松比较 CO₂ 测量结果

CO₂ 浓度值会自动补偿至用户设定的参考温度,确保结果可直接比对、可重现。该智能功能无需在测量前手动调整或调节样品温度,从而节省宝贵的时间并最大限度地减少操作员的工作量,同时保证精度与可靠性稳定一致。  

Gloved hand adjusting temperature correction settings on the touchscreen of the Anton Paar TPO 5001 Total Package Oxygen Meter.

所有的饮料参数:互联、可追溯、数字化

TPO 测量仪系列由多参数饮料分析领域的市场领导者开发,可无缝集成至行业顶尖的多参数检测平台,单一样品即可完成 50 余项产品放行参数分析。全面兼容安东帕实验室执行系统 AP Connect,实现高效数字化数据管理,在生产与实验室网络间提供可追溯、结构化、易调取的测量结果。

Lab technician loading a beer bottle into the Anton Paar TPO 5001; DMA 4501 M and PFD filling device visible in the background.

规格

 TPO 3001TPO 5001
测量范围微量范围传感器宽范围传感器
气相中的氧0 hPa 至 45 hPa 0 hPa 至 1000 hPa
溶解氧0 ppm 至 2 ppm0 ppm 至 45 ppm
CO2[1]30 ℃ (86 °F) 时,0 g/L 至 8 g/L (0 vol. 至 4 vol.) 
<15 ℃ (59 °F) 时,0 g/L 至 10 g/L (0 vol. 至 5 vol.)
温度[2]0 ℃ 至 40 ℃(32 °F 至 104 °F)适用于非冷冻样品
压力5 bar 至 6.2 bar(绝对压力)
重复性[3]
TPO±8 ppb 或 ±6 %,以较大者为准±25 ppb 或 ±6 %,以较大者为准
CO2[1]0.01 g/L (0.005 vol.)
环境条件
环境温度15 ℃ 至 35 ℃(50 °F 至 95 °F),0 ℃ 至 40 ℃(32 °F 至 104 °F)按要求  
相对湿度(无冷凝)[4] 
 
10 % 至 90 %
包装尺寸
包装直径35 mm 至 90 mm(1.4 in 至 3.5 in)
包装高度30 mm 至 370 mm(1.2 in 至 14.6 in)
包装体积[5]150 mL 至 2000 mL
测试包装和样品类型[6]
玻璃(不含软木塞)12 oz、14.5 oz、18 oz、22 oz、22.4 oz、24 oz、32 oz、40 oz | 250 mL、330 mL、750 mL、1000 mL
PET[7]42 oz | 1000 mL、1500 mL
铝瓶14 oz | 2000 mL
铝罐8 oz、10 oz、12 oz、16 oz、25 oz | 330 mL、500 mL
常规样品类型所有常见类型的啤酒、苹果酒、即饮 (RTD) 饮料、苏打水、水、葡萄酒和碳酸非酒精饮料(可应要求提供能量饮料和高碳酸饮料)。所有样品不得含有果肉。
更多详情
耗气量/测量Vn = 13 L
所需的无氧气体N2  或 CO2  5 级(仅限 TPO 3001),5 bar 至 6.2 bar(绝对压力),无波动(±0.1 bar 绝对压力) 
供水[8]- 水质:去离子,无氯,150 µm(100 目)过滤器
- 压力:1.5 bar 至 4.5 bar(绝对压力)
通讯接口2x USB、以太网、CAN(仅供安东帕设备使用)、RS232
电源AC 85 V 至 264 V,47 Hz 至 63 Hz 或 DC 120 V 至 370 V
显示屏7'' LCD 面板触摸屏
数据存储多达 5000 个测量数据集
防护等级IPX3
尺寸(长 x 宽 x 高)515 mm x 590 mm x 1120 mm (20.3 in x 23.2 in x 44.1 in)
重量70 kg (154 lbs)
海拔高度最大 2,000 m

[1] CO2 规格仅适用于:TPO 5001,或者已升级 CO2 的 TPO 3001。
[2] 碳酸化水平 > 5.5 g/L 时 35 ℃ (95 °F),碳酸化水平 > 10 g/L 时 10 ℃ (50 °F)。
[3]在环境温度和样品温度为 23 ℃ (73.4 °F) 且采用标准清洗的情况下。请注意,在确定该组的重复性时,不考虑对该组的第一次测量。
[4] 在较高相对湿度下,需要用干燥空气吹扫仪器。
[5]仅限 TPO 3001;对于 TPO 5001,>300 mL。
[6]根据要求批准其他包装类型。根据要求批准其他样品类型。
[7]仅灌装碳酸样品。
[8]必须对水进行去矿物质处理,以避免钙沉积物(水垢)的积累,防止损坏仪器。

安东帕认证服务

安东帕在服务和支持方面的质量:
  • 全球超过 350 名制造商认证的技术专家
  • 用当地语言提供合格的服务支持
  • 为您的投资提供整个生命周期保护
  • 3 年质保
更多详情

文档