| Required sample volume (mL) |
2.5 |
|
| Minimum solvent consumption (mL per sample) |
5 |
|
| Typical solvent consumption (mL per sample) |
10 |
|
| Volume of the solvent bottles (L) |
2 x 0.5 |
|
|
| Reproducibility |
| |
|
| Viscosity |
0.35 % within the works adjustment range 1 % outside this range |
|
| Density (viscosity compensated) |
0.0005 g/cm³ from 0.65 to 1.5 g/cm³ 0.0020 g/cm³ outside this range |
|
| Temperature |
0.02 °C (0.04 °F) from 15 to 105 °C (59 °F to 221 °F) 0.05 °C (0.09 °F) outside this range |
|
|
| |
|
| Repeatability |
| |
|
| Viscosity |
0.1 % |
|
| Density |
0.0001 g/cm³ |
|
| Temperature |
0.005 °C (0.01 °F) |
|
|
| |
|
| Measuring range |
| |
|
| Dynamic viscosity (mPa.s) |
0.2 to 20 000 |
|
| Max. filling viscosity (mPa.s) |
2000 |
|
| Density (g/cm³) |
0.65 to 3 |
|
| Temperature |
15 to 105 °C (-56 °C with counter cooling at -40 °C) 60 to 220 °F (-68 °F with counter cooling at -40 °F) |
|
|
| |
| Sample throughput (per hour) |
| Values in parenthesis for optional pressed air drying |
| |
|
| Diesel fuel, fast test |
15 (20) |
|
| Engine oil, fast test |
10 (12) |
|
| 1000 mPa.s at 20 °C, precise test |
7.5 (11) |
|
| Engine oil, precise double determination |
5 (6) |
|
|
| |
|
| Data memory/modes |
384 data sets/10 programable modes |
|
| Interfaces |
dual RS-232, AT-keyboard |
|
|
| |
|
| Weigth net/gross |
15/17 kg 33/38 lbs |
|
| Space requirements L x W x H |
440 x 315 x 220 mm 17.3 x 12.4 x 8.7" |
|
| Ambient conditions |
15 to 35 °C (59 to 95 °F), max. 80 % r.h. below 31 °C (88 °F), 67 % at 35 °C (95 °F) |
|
| Power supply (sinusoidal current) |
100 to 240 VAC, 50 to 60 Hz, 75 VA |