| Brix Scale (ICUMSA 1974)168 | ABBEMAT® HP | ABBEMAT® WR |
| Range |
1.32 - 1.56 nD |
1.30 - 1.72 nD |
| Resolution |
1 x 10-6 nD
|
1 x 10-6 nD
|
| Accuracy |
2 x 10-5 nD |
4 x 10-5 nD |
| Brix Scale (ICUMSA 1974)168 |
| Range |
0 - 100 % |
0 - 100 % |
| Resolution |
0.001 % |
0.001 % |
| Accuracy |
0.015 % |
0.030 % |
Automatic Temperature Compensation |
selectable |
selectable |
| |
|
|
| Other RI Derived Scales |
more than 20, continuously extending |
| User Defineable Scales |
linear, polynomial, formula (integrated formula parser) |
| Calibration |
One or two point calibration |
| Materials in Contact with Samples |
| Prism |
YAG (Yttrium Aluminium Garnet) |
| Sample mold |
Stainless steel |
| Seal |
Kalrez™ |
| |
|
| Sample Cover |
Hinged flap with latch and exchangeable sample presser for attaching a flow cell |
| Measuring Principle |
Critical angle of total reflection measured by shadowline detection with 2048-element linear CCD array and digital data processing |
| Methods |
Complies with ICUMSA, ASTM, SSDT, ISI, AOAC, OIML, JIS, DIN, CID methods |
| Light Source |
High performance LED lamp, average life time >100,000 h |
| Wavelength |
589.3 ±0.1 nm (sodium D) by wavelength-adjusted interference filter |
Sample/Prism Temperature Control
|
Built-in solid state thermostat (Peltier), range 10 - 70 °C (ambient temperature 10 - 40 °C), optional 10 - 110 °C, accuracy ±0.03 °C, stability ±0.002 °C |
| Speed of Thermostating |
Approx. 20 seconds for cooling water from room temperature down to 20.00 °C |
Temperature Control of Internal Refractometer Components
|
Peltier solid state thermostat keeps all internal refractometer components at a constant temperature |
| |
|
| Mechanical Data |
| Case Dimensions | W x H x D | Weight |
| Measuring Unit |
Stainless steel |
180 x 120 x 250 mm |
appr. 6 kg |
| Power Supply/Interface Unit |
Anodized aluminum |
135 x 80 x 220 mm |
appr. 2 kg |
| |
|
|
|
| Interfaces |
RS-232C serial port for bidirectional communication with a PC |
| Operator Console |
Personal Computer (PC) required |
| Software |
Set-up and operation in English, German, Russian, Spanish, Portugese, French, Polish, Dutch, other languages upon request |
| Power Requirements |
100 - 240 VAC +10 %/-15 %, 50/60 Hz, min. 10 W, max. 100 W, depending on sample temperature setting and ambient temperature |