Millikelvin Thermometer: MKT 50

In combination with two highly precise Platinium Resistance Thermometers your MKT 50 Millikelvin Thermometer measures the temperature to the highest level of accuracy. MKT 50 is by far the best-priced thermometer for applications which require a low measuring uncertainty of 0.01 °C to 0.001 °C.

Highest accuracy

  • Resolution 0.0001 °C (0.1 mK)
  • Measuring uncertainty over the whole measuring range 0.001 °C (= 1 mK)
  • Anton Paar's patented direct current measuring method (previous models MKT 25/100)

Temperature calculation according to conventional methods

  • DIN IEC 751 for industrial Platinum Resistance Thermometers (PRT)
  • ITS-90 for standard thermometers (SPRT: Standard Platinum Resistance Thermometer)
  • Comparison calibration (PRT) or fixed point calibration (SPRT)
  • Calibration data for 30 sensors can be saved

Highest stability - easy adjustment

  • AD converter does not drift
  • Adjustment of the internal reference resistance by connecting a standard resistor
  • Automatic new calculation of the reference resistance in less than 2 minutes

Modern interfaces plus battery and mains operation

  • LAN connection (Ethernet)
  • Serial interface RS-232
  • Mains adapter
  • Battery operation: 2 AA batteries (>10 hours)

Applications

  • Highly accurate comparison calibrations of all types of thermometers in liquid baths
  • Reference thermometer for all applications and branches
  • Calibration of standard thermometers at temperature fixed points according to ITS-90
Measuring range
Temperature -260 to +962 °C
(-418 to +1764 °F)
Resistance 0 to 440 Ω
Uncertainty of measurement*
Temperature < 1 mK (Pt 100)
Resistance
< 400 µΩ
Resolution
Temperature 0.1 mK (Pt 100)
Resistance 40 µΩ
Measuring sensor Pt 100 or Pt 25.5
(DIN IEC 751 or ITS-90)
Measuring current 0.41 mArms
Internal reference resistor 400 Ω (Vishay)
Number of sensor inputs 2
Sensor connection 4 wire (Lemo 1S304)
Measuring time (for both channels) 1.44 seconds 
Data outputs RS 232 D (optionally USB with USB-RS232 converter); LAN (Ethernet, 10 Mbit)
Ambient operating temperature 0 to 35 °C (for highest accuracy 20 to 25 °C recommended)
Temperature coefficient < 0.3 ppm/°C
Power supply Mains adapter 7.5 V or 2 x AA batteries
Dimensions (L x W x H) approx. 240 x 190 x 110 mm (9.5 x 7.5 x 4.3 inches)
Weight approx. 2 kg (4.4 lbs)

*) refers to the calibration of the reference resistor without contribution of the sensor (confidence level: 95 %, number of measurement values: 50, reference temperature: 23 °C)