Capillary Flow Porometers for Through-Pores:
Porometer 3G

  • +3
  • 360°
Model:
  • Measures pores from 0.07 μm to 100 μm
  • Used with an inlet gas source with up to 100 psi (6.9 bar)

Porometer 3G instruments deliver accurate and repeatable through-pore size measurements within 30 minutes for any sample. Their simplified operation and general measurement speed make these capillary flow porometers the clear choice for through-pore size distribution measurements of filtration media. Depending upon the pore size range and flow-through characteristics of your samples, there’s a Porometer 3G that’s right for you.

Key features

No need for calibrations: saving operator time and effort

No need for calibrations: saving operator time and effort

The pressure sense line in Porometer 3G instruments measures the pressure directly at the sample, and is therefore unaffected by changing flow rates. In contrast, conventional capillary flow porometers measure the pressure at some fixed point prior to applying the sample, and use sample-specific flow path calibrations to predict the pressure at the sample. Because the amount of airflow within this path affects the pressure gradient that exists, each sample type requires a calibration to be performed and applied. The unique design of Porometer 3G eliminates the need for calibrations. This means much less effort and ultimately more accurate measurements.

Automatic pressure regulation ensures ideal measurement conditions every time

Automatic pressure regulation ensures ideal measurement conditions every time

Porometer 3G instruments use an intelligent pressurization routine designed to ensure results within equilibrium throughout the measurement. This routine eliminates the problems that come with conventional systems, due to the fact that measurements made with a constant ramp rate, which is too fast, provide smaller-than-correct pore sizes, as they don’t allow enough time for equilibrium to be achieved at each pressure data point. Conversely, similar measurements made too slowly accentuate the effects of wetting fluid evaporation, providing larger-than-correct pore sizes. Porometer 3G units automatically regulate the pressure - which leads to accurate results independent of the settings. 

Complete through-pore size distribution for the full set of data

Complete through-pore size distribution for the full set of data

Porometer 3G makes all flow measurements on a single flow meter, and thus obtains a continuous account of through-pores present in your material. This includes the crucial region between the maximum and mean pore sizes. Capillary flow porometers that switch measurement devices after the bubble point is reached run the risk of missing a portion of this data. You can rely on Porometer 3G to deliver a complete data set.

Wet-then-dry, dry-then-wet – you choose

Wet-then-dry, dry-then-wet – you choose

Capillary flow porometry measurements involve both a “wet curve” (the flow resulting from the expulsion of wetting fluid from pores at different pressures) and a “dry curve” (the flow through the material over the pressure range of the measurement when wetting fluid is present). Measuring wet-then-dry is much simpler for the operator and does not involve opening the sample holder in the middle of the experiment. Dry-then-wet measurements remove concerns about potential effects on the sample during the creation of the wet curve. Both measurements are available on all Porometer 3G instruments. You can enjoy the workflow ease of wet-then-dry measurements and still have the option to perform dry-then-wet measurements when samples warrant it. 

Adapts to the wide range of sample requirements in your lab

Adapts to the wide range of sample requirements in your lab

All Porometer 3G instruments have the ability to use different shape factors and different fluids in order to develop a test method specifically catered to your sample measurement needs. In addition, their design is perfect for adapting to samples of various sizes and shapes via external holders for use with hollow fiber, solid piece, or rigid tubular configurations. 

Technical specifications

  3G micro 3G z 3G zH
Pore size, minimum 0.07 µm <0.03 µm <0.02 µm
Pore size, maximum 100 µm 500 µm 500 µm
Pressure controllers 1 2 2
Controller #1 0 psi to 100 psi 0 psi to 30 psi 0 psi to 30 psi
Controller #2 n/a 0 psi to 300 psi 0 psi to 500 psi
Pressure sensors 2 3 3
Sensor #1 0 psi to 5 psi
Sensor #2 0 psi to 100 psi
Flow sensors 1 1 2
Sensor #1 0 L/min to 100 L/min 0 L/min to 100 L/min (standard) 0 L/min to 10 L/min
Sensor #2 n/a n/a 0 L/min to 200 L/min
Flow sensor switching n/a manual auto
Pressure sensing accuracy ±0.05 % f.s.
Pressure resolution 16 bit A/D equivalent
Flow sensor type Precision thermal mass flow
Flow sensor settling time <2 seconds
Flow sensor temperature coefficient <0.05 %/°C (15 °C to 45 °C)
  General specifications
Electrical properties 90 V to 240 V AC, 50/60 Hz
Main control unit dimensions (H x D x W) 40 cm x 45 cm x 19 cm (15.8 in x 17.8 in x 19 in)
Sample holder block dimensions (H x D x W) 24 cm x 32 cm x 15 cm (9.5 in x 12.6 in x 6 in)
This table shows standard configurations. Different pressure and flow rate ranges are available for some models.

Anton Paar Certified Service

The Anton Paar quality in service and support:
  • More than 350 manufacturer-certified technical experts worldwide
  • Qualified support in your local language
  • Protection for your investment throughout its lifecycle
  • 3-year warranty
Learn more

Documents

Categories
Industry
Application
Product
Your choice: Reset all filters

Consumables

Consumables

Not all items are purchasable online in specific countries.

To find out if you can purchase online from your location, check the online availability below.

3G TOP SUPPORT SCREEN FOR 50 mm HOLDER

Compatible with:
Porometer 3G micro | z | zH
Online price only | excl.incl. 0VAT
Delivery time: %1$s – %2$s working days %1$s – %2$s weeks %1$s – %2$s months
Part number: 197697