| | Alcolyzer 5001 | Alcolyzer 7001 |
| Alcohol |
| Measuring range | | |
| - Beer products | 0 to 12 % v/v |
| - Wine products | 0 to 20 %v/v |
| - Sake products | 0 to 20 %v/v |
| - Whiskey products | 35 to 65 %v/v |
| - Cognac products | 0 to 65% v/v |
| - Vinegar products | - | on request |
| Typical deviation to distillation (all products) | < 0.1%v/v |
| Repeatability s.d. | 0.03 %v/v | 0.01 %v/v |
| Color (optional) |
| Measuring range | - | 0 to 120 EBC (0 - 60.96 ASBC) |
| Repeatability s.d. | - | 0.1 EBC (0.05 ASBC) |
| Temperature |
| Displayed temperature 1) | 15 °C or 20 °C (59 °F or 68 °F) |
| Repeatability s.d. | 0.01 °C (0.018 °F) |
| Sample temperature | 18 to 25 °C (64.4 °F to 77 °F) |
| Typical measuring time | 1 to 3.5 min |
| Sample volume | 20 mL (syringe filling) vs. 40 mL (sampler filling) |
| Typical samples | | |
| - Beer products 2) | Beer, non-alcoholic beer, beer mixed drinks, RTDs, molasses, wash, kombucha, hard seltzer, 0.0% beers on request |
| - Wine products 2) | Sake, cider, wine, fortified wine, non-alcoholic wines on request |
| - Sake products 2) | Sake, cider, wine, fortified wine, non-alcoholic wines on request |
| - Whiskey products 2) | Whiskey, tequila, spirits with < 5 g/L extract |
| - Cognac products 2) | Cognac, rum, spirits with 5 - 20 g/L extract, non-alcoholic spirits on request |
| | Vinegar on request 4) |
| Instrument data and operating conditions |
| Data memory | Measurement data: up to 10,000 Check results: up to 300 Adjustment results: up to 300 Products: up to 40 Custom quantities: up to 50 Images: up to 50 |
| Display | 10.1" TFT WXGA (1280x800 px); PCAP touchscreen |
| Controls | Touchscreen, optional keyboard, mouse, and barcode reader |
| Interfaces | 4 x USB 2.0 (type A)3), 1 x USB OTG (Micro-A), 1 x RS-232, 1 x Ethernet (100 Mbit), 1 x CAN Bus |
| RS-232C printer settings | Baud rate: 9600; Parity: none; Stop bit: 1; Data bits: 8 |
| Environmental conditions (EN 61010) | Indoor use only |
| Ambient temperature | +18 °C to +32 °C (+59 °F to 89.6 °F) |
| Air humidity | Non-condensing 20 °C: < 90 % relative humidity 25 °C: < 60 % relative humidity 30 °C: < 45 % relative humidity |
| Altitude | Max. 3000 m (9800 ft) |
| Maximum operating pressure | Ambient pressure |
| Pollution degree | 2 |
| Overvoltage category | II |
| Voltage | 100 to 240 V~, 50/60 Hz, fluctuation ±10 % |
| Power consumption | 90 VA |
| Power inlet | According to IEC/EN 60320-1/C14, protection class I |
| Fuses | Ceramic tube fuses 5x20 mm; IEC60127-2; AC 250 V; T 5 AH |
| Dimensions (L x W x H) | 526 mm x 347 mm x 230 mm (20.7 in x 13.7 in x 9 in) |
| Weight | 17.6 kg |