Microwave digestion transforms your sample preparation into a faster, safer, and more reliable process. Compared to conventional hot block or hot plate methods, digestion times are up to 20x faster, significantly increasing throughput.
The closed-vessel system minimizes acid exposure and lowers consumption while ensuring precise temperature and pressure control. Cleaner digests improve the LOD and LOQ performance of your ICP, giving you greater confidence in every reported value.

Book your demo

Why switch to microwave digestion?

In open-vessel systems such as hot plates or hot blocks, digestion performance is limited by the boiling point of the acid mixture. Higher temperatures accelerate reactions but also cause evaporation losses. Closed-vessel microwave digestion allows higher temperatures and pressures, prevents contamination and acid loss, and delivers faster, safer, and more reliable results.

Minimalist clock icon with white circular outline and two hands on dark background, symbolizing time saving or duration

Faster

Higher temperatures under pressure accelerate reaction kinetics and significantly reduce digestion time compared to hot blocks or hot plates. Shorter preparation times mean lower labor costs per sample and faster reporting, turning saved minutes into measurable financial value.

Dark square icon with rounded corners and subtle gradient, used as a generic placeholder or background element

Safer

Closed-vessel systems prevent acid evaporation and minimize operator exposure to corrosive fumes, while lower acid consumption and controlled pressure conditions enhance laboratory safety and compliance.

Dark rounded square with subtle gradient, minimal UI icon background or placeholder element

More reliable

Complete digestion in sealed vessels minimizes contamination and background interference, improving the LOD and LOQ performance of your ICP. Higher reproducibility reduces measurement deviations and significantly lowers the need for costly and time-consuming remeasurements.

Assorted food samples on white surface: minced meat, coconut pieces, fat, peas, potato, soybeans, oil, milk, and white powder in dishes

Food analysis: Safe and compliant results

Food laboratories face strict regulatory limits, complex matrices, and high sample volumes. Incomplete digestion can compromise trace element analysis and delay batch release. Microwave digestion ensures fast, complete mineralization of diverse food samples, delivers the lowest blank values (and very low detection limits), and improves reproducibility. Higher throughput and reliable ICP performance support compliance, protect brand integrity, and accelerate time to market.

Learn more

Laboratory dishes with different soil samples: clumps, fine powder, sand, organic matter, and wood chips on white background

Environmental: Accurate trace monitoring

Environmental labs must detect trace elements at very low concentration levels in soil, water, sludge, and waste. Complex matrices and contamination risks can affect LOD and LOQ performance. Closed-vessel microwave digestion minimizes blank values, prevents analyte loss, and ensures complete digestion. The result is higher data quality, improved sensitivity, and compliance with environmental regulations.

Learn more

Gloved hand using tweezers to pick round pellet from dish, with containers of animal feed pellets and cubes on white surface

Feed testing: Reliable nutrient control

Feed analysis requires accurate determination of minerals and trace elements to ensure animal health and regulatory compliance. However, heterogeneous matrices and high fat content can be challenging for conventional digestion methods. Microwave digestion delivers controlled high-temperature conditions for complete sample breakdown, improving recovery rates and reducing remeasurements. Faster workflows increase laboratory efficiency and ensure consistent feed quality.

Lear more

Top-down view of capsules, cream jar, petri dish with gel, syringe, dropper, test strips, and wooden sticks on white surface

Cosmetics and pharmaceuticals: Precision and compliance

Cosmetic and pharmaceutical products require strict impurity control, full regulatory compliance, and complete traceability. Microwave digestion enables controlled high-temperature reactions, ensures complete mineralization, and delivers reproducible results. Improved ICP performance, individual sample documentation for process control, and robust data integrity support compliant workflows and confident batch release.

Learn more

From Open Digestion to Controlled Performance

Reliable digestion for complex environmental samples

EPA 3050B and EPA 3051A are widely used methods for acid digestion of environmental samples prior to elemental analysis. While EPA 3050B relies on open vessel hot block digestion, EPA 3051A uses closed-vessel microwave technology to achieve higher temperatures and controlled conditions.

ParameterHot block
(EPA 3050B)
Microwave
(EPA 3051A)
Advantage 
Temperature95 °C175 °CHigher temperature enables more complete digestion, fewer residues, and improved recovery
Heating time2 h to 6 h10 minClearly defined method endpoint ensures reproducible results, avoids operator-dependent reheating, and reduces overall turnaround time
Sample weight1 g to 2 g250 mg to 500 mgSmaller sample weight reduces matrix effects and improves digestion consistency
Acid consumption10 mL to 30 mL HNO3 + 3 mL to 10 mL H2O210 mL HNO3 or 9 mL HNO3 + 3 mL HClFixed acid volumes and closed conditions reduce blank values and eliminate repeated acid additions, improving consistency and detection limits

Microwave digestion delivers faster, safer, and more reliable sample preparation. Clearly defined methods, controlled conditions, and reduced acid consumption eliminate operator-dependent variability, minimize rework and ensure consistent, high-quality results across all sample types.

Faster and controlled preparation of aqueous samples

Standard Method 3030E, EPA 3010A, and EPA 3015A define standard approaches for preparing aqueous samples for elemental analysis. Traditional hot block methods such as 3030E and 3010A rely on open digestion with manual intervention, whereas 3015A applies microwave technology to enable controlled, high-temperature digestion in closed vessels, ensuring no contamination and no loss of analytes.

ParameterHot plate
(SM 3030 E)
Hot plate
(EPA 3010A)
Microwave
(EPA 3015A)
Advantage
Temperature~ 100 °C90 °C to 95 °C170 °CElevated temperature improves digestion efficiency and ensures complete element recovery
Heating time> 2 h> 2 h20 minShort, program-controlled runs significantly reduce time per sample,  enabling faster result delivery and more efficient use of lab resources
Sample volume100 mL100 mL22.5 mLBalanced sample volume supports efficient reactions while maintaining analytical accuracy
Acid consumption5 mL HNO6 mL to 15 mL HNO3 + ~ 2 mL HCl2.5 mL HNO3 or 2 mL HNO3 + 0.5 mL HClDefined reagent amounts avoid manual adjustments and reduce variability between runs

Microwave-based digestion streamlines sample preparation by shortening processing times, improving lab safety through closed systems and delivering consistent, reproducible outcomes. Clearly defined parameters and superior digestion performance bring clarity to laboratory workflows, eliminating questionable results and reducing the need for remeasurements while supporting reliable analytical data across all sample matrices.