FTIR Oil Analysis:
Lyza 7000 Used Oil Package
- Quick Oil Condition Monitoring (OCM) using FTIR spectroscopy in under 45 seconds
- Standards-compliant analysis according to ASTM E2412 and more
- In-service lubricant FTIR analysis out-of-the box
- Easy spectral characterization, even for less experienced users, with one press of a button
Perform oil analysis at a touch, using Anton Paar’s FTIR. Preinstalled methods and corresponding transmission cells ensure compliance with ASTM standards. The software seamlessly integrates measurement, processing, and spectral analysis. Automated workflows and customizable reporting make your oil condition monitoring ultra-efficient.
Key features
ASTM-compliant FTIR oil analysis
Analyze your spectra directly in the instrument’s embedded software.
Lyza 7000 ensures compliance with ASTM E2412, D7214, D7412, D7414, D7415, D7624, D7844, and D7418 via pre-installed methods.
Measure parameters like soot, oxidation, nitration, phosphate anti-wear components, sulfate by-products and the contamination of water, gasoline, diesel, and ethylene glycol quickly and reliably.

Everything you need for your FTIR oil analysis
Lyza 7000 is a compact benchtop spectrometer for used oil analysis, with an adjustable touchscreen that works right out-of-the-box.
The transmission module is equipped with a 2 in x 3 in card holder, supporting industry-standard transmission cells. The 100 µm cell for liquids, suited for ASTM-compliant measurement, is a perfect fit.
Efficient on every level, Lyza 7000 is packed with convenience-maximizing features: guided workflows, continuous instrument self-monitoring, and energy-saving mode. Software tailored to condition monitoring of in-service lubricants rounds off a perfect package.

The FTIR spectrometer you can trust
Lyza 7000 is a spectrometer you can count on for FTIR oil analysis.
Rigorous testing and validation underpin consistent, accurate results. Constant internal system monitoring ensures potential issues can be resolved quickly. A smart status LED light updates you on measuring progress at a glance.
This FTIR spectrometer was designed by pioneers in optical analysis – and it shows!

FTIR oil analysis with minimum costs and maximum efficiency
The Lyza 7000 software enables fast analysis of key parameters for used oil analysis by even inexperienced operators. An automatic polystyrene check with the internal reference occurs at the push of a button and takes just minutes.
The touchscreen, embedded software, and small footprint save bench space. And programmable “Eco Mode” reduces energy use and prolongs the lifetime of core components.
Lyza 7000 offers cost efficiency with zero compromise on quality.

Specifications
Detector | Pyroelectric DLaTGS detector |
Optics | Hermetically sealed aluminum casing with gold-coated mirrors, KBr windows, and beam splitter |
Signal-to-noise ratio | 55,000:1 (1 min, 8 cm-1, 1,900 cm-1 to 2,250 cm-1) |
Spectral range | 350 cm-1 to 7,500 cm-1 |
Spectral resolution | 1.4 cm-1 to 16 cm-1 |
Wavenumber accuracy | <0.05 cm-1 @ 900 cm-1 to 3,000 cm-1 |
Measurement time | <30 seconds |
Laser type | Single mode Vertical Cavity Surface Emitting Laser (VCSEL) |
Laser class | Class 1, enclosed hermetically |
IR source | SiC composite |
Interferometer | Permanently aligned cube corner interferometer |
Desiccant | Molecular sieve with color indicator |
Operating temperature range | 10 °C to 30 °C |
Dimensions (D x W x H) | 365 x 315 x 382 mm |
Weight | 12.8 kg |
Power supply | DC 24V, 45 W |
Communication interfaces | 4 x USB 2.0 / CAN / Ethernet / Wi-Fi* |
Data export formats | .csv, .pdf, .spc |
Display | 10.1” PCAP touchscreen, multi-touch |
Controls | Touchscreen, optional keyboard, mouse, and barcode reader |
Internal storage | 32 GB |
*via external Wi-Fi stick
Parameter | Standard | Unit |
Degradation | ||
Oxidation | ASTM E2412 | Abs/0.1mm Abs/cm |
D7214 | Abs/mm | |
D7414 | Abs/mm | |
Nitration | ASTM E2412 | Abs/0.1mm |
D7624 | Abs/cm | |
Sulfate by-products | ASTM E2412 | Abs/0.1mm |
D7415 | Abs/cm | |
Additives | ||
Antiwear Components (ZDDP) | ASTM E2412 | Abs/0.1mm |
D7412 | Abs/cm | |
Contaminants | ||
Soot | ASTM E2412 | Abs/0.1mm |
D7844 | % | |
Water | ASTM E2412 | Abs/0.1mm |
Ethylene Glycol (Antifreeze) | ASTM E2412 | Abs/0.1mm |
Diesel Fuel | ASTM E2412 | Abs/0.1mm |
Gasoline | ASTM E2412 | Abs/0.1mm |
Standards
ASTM
Anton Paar Certified Service
- More than 350 manufacturer-certified technical experts worldwide
- Qualified support in your local language
- Protection for your investment throughout its lifecycle
- 3-year warranty
Documents
Similar products
Consumables & Accessories
Consumables & Accessories
If you do not find the item you require, please contact your Anton Paar sales representative.
To find out if you can purchase online from your location, check the online availability below.

POWER SUPPLY 24V 3A 72W

KEYBOARD GERMAN USB

KEYBOARD USA USB

PROTECTING COVER FOR KEYBOARD

BARCODE SCANNER 2D WITH USB CABLE
Only for ViscoQC 300

ETHERNET CABLE FOR PC CONNECTION CAT5E, RJ45G-RJ45G, SFTP, 2 m, CROSSED
Only for ViscoQC 300. Cable for LIMS Bridge connection to PC for transferring measurement data.

WIFI/BLUETOOTH DONGLE