| | Diana 700 | Diana 300 |
| Standard test methods | ASTM D86 (Group 0, 1, 2, 3, 4), ASTM D850, ASTM D1078, EN ISO 3405 (Group 0, 1, 2, 3, 4), IP 123, IP 195, JIS K 2254, GOST 2177, ASTM D524, ASTM D4530 | ASTM D86 (Group 0, 1, 2, 3, 4), EN ISO 3405 (Group 0, 1, 2, 3, 4), IP 123, JIS K 2254, GOST 2177 |
| Operation |
| User interface | 10" TFT color touchscreen, solvent-proof |
| Heating system | - Low mass/low voltage heating system with automatic positioning of heater and safety shield
- Automatic initial heater settings and heater regulation
- Cooling fan to quickly reach safe handling temperature after test
| - Low mass/low voltage heating system with manual positioning of heater and safety shield
- Preprogrammed initial heater settings in test methods
|
| Condenser system | - Liquid-free cooling with Peltier technology
- Temperature range: 0 °C to 80 °C, resolution: 0.1 °C
|
| Receiving chamber | - Liquid-free cooling with Peltier technology
- Temperature range: 0 °C to 80 °C, resolution: 0.1 °C
- Corrosion-proof
- Automated reduced VOC emission
|
| Vapor temperature | Pt100, class A (metal or glass) - Temperature range: 0 °C to 450 °C, resolution: 0.1 °C
- Integrated calibration memory with 20 calibration points and automatic read-out (ID detection)
- Automatic correction by barometric pressure
| Pt100, class A (glass) - Temperature range: 0 °C to 450 °C, resolution: 0.1 °C
- Automatic correction by barometric pressure
|
| Sample volume | Optical static detection system with contact image sensor (CIS) technology |
- Sample volume scan before test start to measure initial volume (converted to 100 %, if necessary)
- Residue scan in receiving chamber with automatic calculation of loss
- Volume range: 0 mL to 103.5 mL, resolution: 0.01 mL
- Accuracy: ±0.1 mL
| - Volume range: 0 mL to 100 mL, resolution: 0.01 mL
- Accuracy: ±0.1 mL
|
| Ambient pressure | Built-in pressure sensor, range: 30 kPa to 110 kPa, resolution: 0.1 kPa |
| Ambient humidity | Built-in humidity sensor, range: 0 % to 95 % (rel. humidity) |
| Ambient temperature | Built-in temperature sensor, temperature range: -20 °C to +80 °C, resolution: 0.1 °C |
| Status light | Yes | No |
| Test results | >4000 | 200 |
| User defined test methods | >200 | 15 |
| Safety |
| Fire extinguisher | - Built-in fire extinguisher with IR sensor for fire detection
- Connection of 4 bar to 6 bar CO2 or N2
|
| Condition monitoring system | Detection of flask size, flask support board, vapor temperature sensor, heater shield, heater lift, condenser cleaning, receiving chamber door, receiving cylinder, drip plate, detection if fire extinguishing gas is connected | Detection of vapor temperature sensor, receiving chamber door, receiving cylinder, drip plate |
| Operating Requirements |
| Temperature | 10 °C to 35 °C |
| Relative humidity | 10 % to 80 % relative humidity at 35 °C |
| Voltage | 100 V to 240 V, 50/60 Hz |
| Total power | 900 W |
| Altitude | 0 m to 5000 m |
| Dimensions and Weight |
| Dimensions | approx. 406 mm × 525 mm × 645 mm (W × D × H) |
| Weight | approx. 40 kg |
| Further Features |
| Connectivity | 1 × Ethernet, 4 × USB, Bluetooth, Wi-Fi, AP Connect, LIMS | 2 × USB |
| Automatic calculation | Cetane index, driveability index | |
| Accessories | Dry point sensor, 200 mL flask, bar code reader, printer, receipt printer, keyboard, various certified reference materials (CRM) | bar code reader, printer, receipt printer, keyboard, various certified reference materials (CRM) |