| | TAG 300 | TAG 500 |
| Test methods | ASTM D56 Equilibrium procedures: ASTM D3941, ASTM D3934, ISO 1516, ISO 1523 >1,000 user-defined test programs |
| Application range (flash point temperature) | +10 °C to +110 °C | Internal cooling +10 °C to +130 °C External cooling -30 °C to +130 °C |
| Sample temperature range | -7 °C to +110 °C | Internal cooling -7 °C to 130 °C External cooling -35 °C to +130 °C |
| Ignition type | Electric (encapsulated hot wire) | Electric (encapsulated hot wire) Optional gas ignition |
| Heating rate | According to standard or user-defined |
| Cooling | Fan-supported internal Peltier cooling technology | Fan-supported internal Peltier cooling technology or external cooler |
| Barometric pressure correction | Flash point is automatically corrected to barometric pressure |
| Flash detection | Automatic detection by thermocouple |
| Sample temperature | Measured by Pt100 |
| Safety | Overheat protection, automatic shut-off | Detects a “flash” outside the cup Optional automatic fire-extinguishing system in combination with a unique optical fire detection system User management with different access levels | Overheat protection, automatic shut-off | Detects a “flash” outside the cup Automatic fire-extinguishing system in combination with a unique optical fire detection system as standard User management with different access levels |
| Calibration | Calibration of sample and block temperature, and internal barometer |
| Handling | TFT touchscreen | Small footprint | Manual lifting/lowering of multi-function head | No open cables | User-definable measurement parameters and units | Display of test progress in real time | TFT touchscreen | Small footprint | Fully automatic multi-function head | No open cables | User-definable measurement parameters and units | Display of test progress in real time |
| Memory | 1 GB space for approx. 50,000 tests and 1,000 users |
| Statistics | Mean, min, max, repeatability, standard deviation, max. deviation |
| Interfaces | 2 × USB, 1 × LAN | 4 × USB, 1 × LAN |
| Input options | Optional USB keyboard/mouse/barcode reader |
| Display | 7" TFT, PCAP touchscreen |
| Power supply | 100 V to 240 V, 50/60 Hz, 300 W |
| Gas supply | Optional fire extinguisher: CO2 or N2 inert gas; inlet pressure 400 kPa to 500 kPa | Integrated fire extinguisher: CO2 or N2 inert gas; inlet pressure 400 kPa to 500 kPa Optional gas ignition: 50 mbar of propane or butane |
| Dimensions | 262 mm × 472 mm × 437 mm (W × D × H) | 262 mm × 497 mm × 477 mm (W × D × H) |
| Weight | Ca. 14 kg | Ca. 15 kg |
| Options | Standard test insert in brass or stainless steel, calibration set, condensation protection, draft shield Fire extinguisher | Standard test insert in brass or stainless steel, calibration set, condensation protection, draft shield Gas igniter |