| | Diana 701 | Diana 301 |
| Standard test methods |
| Fuel distillation | ASTM D86 (groups 0 to 4), EN ISO 3405 (groups 0 to 4), IP 123, JIS K 2254, GOST 2177 |
| Solvent distillation | ASTM D1078, ASTM D850, IP 195, JIS K 0066 | – |
| Additional methods | ASTM D524, ASTM D4530, Benzinum Medicinale | – |
| Measurement |
| Vapor temperature sensor | Pt100, metal or glass, Class A (DIN EN 60751) | Pt100, glass, Class A (DIN EN 60751) |
| Vapor temperature range | 0 °C to 450 °C |
| Vapor temperature calibration | Integrated sensor memory for up to 12 calibration points; automatic probe ID detection | Up to 3 calibration points stored on the instrument |
| Calibration certificate | Factory calibration certificate included (room temperature, 100 °C, 180 °C); ISO 17025 accredited calibration available optionally |
| Distillation rate | Up to 10 mL/min, programmable |
| Volume measurement | Optical static detection system with contact image sensor (CIS) and anti-condensation technology |
| Volume range | 0 mL to 103.5 mL | 0 mL to 100 mL |
| Volume accuracy | ±0.1 mL |
| Ambient pressure | Built-in pressure sensor (range: 30 kPa to 110 kPa) with single-point calibration against an external barometer |
| Ambient temperature | Built-in temperature sensor (range: -20 °C to +80 °C) |
| Temperature control |
| Heating system | Low-mass, low-voltage heating system with stabilized heating coil |
| Initial heating control | Automatic | Assisted |
| Condenser and receiving chamber | Liquid-free Peltier (range: 0 °C to 80 °C; accuracy: ±0.5 °C at 25 °C) |
| Operation and safety |
| User interface | 10" solvent-resistant touchscreen |
| Guided mode and status light | Yes | – |
| Automatic volume handling | Initial volume scan with automatic normalization to 100 % and residue scan with automatic loss calculation | – |
| Automatic calculations | Cetane index, drivability index | – |
| Condition monitoring | Advanced setup and safety monitoring, including flask size, support board, vapor temperature sensor, heater shield and lift, condenser cleaning status, receiving chamber door, receiving cylinder, drip plate, and extinguisher gas connection | Essential safety monitoring, including vapor temperature sensor, receiving chamber door, receiving cylinder, and drip plate |
| Heater handling | Automatic heater and shield lift with cooling fan | Manual heater and shield lift |
| Fire safety | Built-in IR fire detection and fire extinguisher using N₂ or CO₂ |
| VOC emission reduction | Automatic |
| Data and connectivity |
| Stored test results | >4,000 | 200 |
| User-defined test methods | >200 | 15 |
| Connectivity | AP Connect, LIMS, network share, email | AP Connect |
| Operating requirements |
| Operating temperature | 10 °C to 35 °C |
| Altitude | 0 m to 2,000 m |
| Power supply | AC 100 to 240 V, 50/60 Hz, 900 VA |
| Technical data |
| Dimensions (W × D × H) | 406 mm × 525 mm × 645 mm |
| Weight | 40 kg |
| Communication interfaces | Ethernet, 4 × USB, Wi-Fi | Ethernet, 2 × USB |
| Dry-point detection | Optional sensor for 125 mL and 200 mL flasks | – |
| Additional accessories | Quartz glass flasks, external printer, receipt printer, barcode scanner, calibration sets, certified reference materials (CRMs) |