Pensky-Martens flash point tester:
PMA 5
- Automatic Pensky Martens closed-cup flash point tester compliant with ASTM D93, ISO 2719, and more
- Maximum safety: integrated fire-extinguishing system
- Pre-installed standard methods, 15 user-defined programs
upon request
The automatic Pensky-Martens closed-cup tester measures the flash point at the lowest temperature at which the application of an ignition source causes the vapors of a sample to ignite.
PMA 5 is suitable for flammability applications on both biodiesel and biodiesel-blended fuels along with distillate fuels like diesel, heating oil, and kerosene. It is also commonly used for other potentially flammable liquids.
Key features
Benefits at a glance
- Integrated fire-extinguishing system automatically controlled or manually released (requires N2 or CO2 gas)
- The One-Twist swivel-around multi-function head guarantees the safe and easy one-hand connection of all sensors and actuators such as the shutter release, stirrer, igniter, and detector contacts.
- A multi-detector combines the flash point detector and the temperature probe in a solid housing with a perfectly adjusted immersion depth
- Memory for 1000 tests, 20 operators, 100 sample names
- Statistical analysis (Min, Max, Mean, Repeatability)
- Data handling via USB stick for import into MS Excel® or to LIMS
Convenient Operation
- Test initiation by pre-installed standard test methods and the option to customize methods in 15 user-defined programs
- Calibration protocol printable
- Menu navigation by jog shuttle system with easy turn-and-push operation and color-coded LED indication for each operation mode
- 5.7” color graphics display shows real-time measurement
- Password protection
Customized user flexibility
- PC-software FPPNet (optional) for convenient remote handling and test data storage
- Stainless steel or small-volume sample cup (optional)
- Stainless steel sample temperature sensor Pt100
- Bar code reader or keyboard (optional)
Technical specifications
Application range | Up to 405 °C (°C/°F selectable) |
---|---|
Ignition | Gas and electric (interchangeable) |
Heating | According to method or user-defined |
Stirring | According to method or user-defined |
Cooling | Built-in forced air (fan) |
Flash point detection | Thermocouple |
Sample temperature | Pt100 glass probe |
Sample temperature calibration | Dynamic calibration against a certified digital thermometer and offset capability with correction table (21 calibration points) |
Barometric pressure correction | Automatic correction |
Safety | Overheat protection, automatic shut-off, fire-extinguishing system, password |
Interfaces | 3x USB, 1x RS232, 1x LAN |
Gas supply | Test flame: 50 mbar of propane or butane Fire extinguisher: 6 bar to 12 bar of nitrogen or carbon dioxide |
Power supply | 115 V/230 V, 50 Hz/60 Hz, 1000 W |
Dimensions | 230 mm x 410 mm x 460 mm (W x D x H) |
Weight | 14 kg |
Resolution | 0.1 °C |
Standards
ABNT
ANP
ASTM
IP
ISO
JIS
RANP
Anton Paar Certified Service
- More than 350 manufacturer-certified technical experts worldwide
- Qualified support in your local language
- Protection for your investment throughout its lifecycle
- 3-year warranty
Documents
Accessories & Software
Consumables
To find out if you can purchase online from your location, check the online availability below.
Check online availabilityIf you do not find the item you require, please contact your Anton Paar sales representative.

Drip tray, 20 ml, aluminum, pack of 10

Lid 2 PMA 4 + 5, PM 4

Multi detector (2) PMA 4 + 5 stainless

Multi detector (2) PMA 4 + 5, glass

Multi detector (2) PMA 4+5 short stainl.

Plug-in connector NW7,2 243.07-B brass

Stopper for multi-detector opening

Storage set for cup, cover + sensor

Test insert PMA 4 + 5, PM 4 complete

Tong