Distillation unit for fuels and solvents

Diana 700 is the ideal solution for automatically performing high-precision distillation range analysis at atmospheric pressure. Typical samples to be characterized include petrochemical products, aromatic hydrocarbons as well as other volatile organic liquids.

The combination of premium technology and distinguished features leads to unrivaled ease of use, high efficiency, and safe operation.

Get in touch Document Finder

Key Features

Smart operator assistance

Smart operator assistance

Diana 700 automates manual handling steps during the measurement setup to provide maximum support. Additionally, for untrained users the innovative user interface employs illustrations as a step-by-step guide through the procedure that is necessary to start a distillation. For experienced users, the advanced mode just shows the state of each component of the instrument and enables a fast start of the distillation. The intelligent condition monitoring system controls all steps and Diana 700 only allows you to start a measurement if all steps have been completed correctly. This helps to avoid having to repeat a measurement due to an incorrect setup or non-compliance with the standard.

Convenient handling: Mobile multi-plug

Convenient handling: Mobile multi-plug

For a perfect operator experience and to increase the usability of the instrument, the vapor temperature sensor was combined with the flask holder. The resulting multi-plug was developed by considering valuable customer feedback. With just one hand and within seconds it is possible to install the distillation flask (125 mL or 200 mL) in the correct position.

The integrated data memory stores the calibration data of the temperature sensor which is automatically applied when the sensor is mounted to the instrument.

Safety first: Automatic heater and shield lift

Safety first: Automatic heater and shield lift

As soon as the distillation flask has been installed the heater and the protection shield move up fully automatically. The correct flask and flask support board are identified by the instrument and only if the correct type is installed can the distillation be started.

For a maximum level of safety, the automatic shield lift surrounds the flask, reducing the risk of touching any hot parts while a measurement is running. After finishing the test, the heater moves down to accelerate the cooling of the flask. This means you save time and can start your next measurement right away.

Easy to operate and efficient: The receiving chamber

Easy to operate and efficient: The receiving chamber

You can operate the receiving chamber of Diana 700 smoothly to access the receiving cylinder. Thanks to the tight jacketing of the cylinder, the chamber guarantees an excellent stability of the sample and cylinder temperature.

Using the highly efficient Peltier technology, Diana 700 enables a fast switch between different ASTM D86 distillation groups and reaches the required temperatures in less than five minutes.

A unique cold trap avoids condensation caused by high humidity and reduces evaporation losses of volatile organic compounds (VOC) at the same time.

Highly accurate volume detector

Highly accurate volume detector

The volume detection of the atmospheric distillation unit Diana 700 makes use of the high accuracy of contact imaging sensor technology (CIS) to achieve a precise determination of the sample volume. This solid state technology is able to continuously follow the liquid level in the cylinder. Additionally, the volume is measured by the instrument before the test starts and corrected to 100 % volume, if necessary. 

Technical Specifications

Tempering system Liquid-free Peltier cooling technology
Temperature range up to 450 °C vapor temperature
0 °C to 80 °C condenser temperature
0 °C to 60 °C chamber temperature
Measuring system Pt100 vapor temperature sensor, class A
Volume detection by contact imaging sensor technology
Safety
  • Automated fire-extinguishing system with detection if gas is connected
  • Automatic self-check at instrument start
  • Intelligent condition monitoring system to avoid incorrect setups
  • Reduced volatile organic compound (VOC) emission
Interfaces USB, LAN, HDMI
Display 10" TFT, PCAP touchscreen
Gas supply N2 or CO2 (6 bar to 12 bar)
Power supply 90 V to 240 V, 50 Hz/60 Hz, 900 W
Dimensions 406 mm x 525 mm x 645 mm (W x D x H)
Weight approx. 40 kg
Other special functions
  • Mobile multi-plug with integrated vapor temperature sensor for wireless flask handling
  • Built-in sensor for ambient temperature, pressure, and humidity

Standards

A

ASTM D86
ASTM D1078
ASTM D850
ASTM D524
ASTM D4530

D

DIN 51751

G

GOST 2177

I

IP 195
IP 123
ISO 3405