Analysis of regular soft drinks with PBA-S Generation M

The PBA-S Generation M (Packaged Beverage Analyzer for Soft Drinks) is the up-to-date measuring system for beverage analysis. Determine sugar (°Brix) and CO2 in soft drinks without prior sample degassing. The beverage is filled into the analysis system directly from its container and is then measured within three minutes: °Brix and CO2 content are determined in a single work step.The modular system consists of the DMA™ 4100/4500/5000 M density meter, the CarboQC ME CO2 measuring module, and the PFD Filling Device (Plus).

Get in touch Document Finder

Key Features

Precise beverage analysis without sample preparation

  • No sample preparation required (no degassing, no filtering)
  • Measurement directly out of glass bottles, PET bottles and cans
  • Precise determination of °Brix and CO2
  • Additional output: Package pressure and dissolved air

Fast, comfortable and safe

  • Measuring time: 3 minutes
  • Modular design (Plug and Play)
  • Intuitive touchscreen operation
  • U-view™: Live image of DMA™ M measuring cell from a built-in camera
  • Safe and clean sample handling with the PFD Filling Device (Plus)

10.4" touchscreen

  • Applies state-of-the-art PCAP technology
  • Provides a premium user experience and overview

Technical Specifications

Technical Data of PBA-S Generation M


Measuring range Repeatability, s.d.
Density 0–3 g/cm³ DMA™ 4100 M: 0.00005 g/cm³
DMA™ 4500 M: 0.00001 g/cm³
DMA™ 5000 M: 0.000001 g/cm³
Concentration sugar 0–15 °Brix DMA™ 4100 M: 0.015 °Brix
DMA™ 4500 M: 0.01 °Brix
DMA™ 5000 M: < 0.01 °Brix
CO2 0–6 vol. (0–12 g/L) at 30 °C (86 °F)
0–10 vol. (0–20 g/L) < 15 °C (59 °F)
0.005 vol. (0.01 g/L)
O2 0–4 ppm 2 ppb (in the range < 200 ppb)
pH pH 0–14 0.02 (in the range pH 3–7)



Typical measuring time 2–3 minutes
Maximum allowed volume of packages to be handled by PFD/PFD Plus Can: 0.5 L
Glass bottle: 1 L
PET bottle: 3 L
Sample temperature 4–25 °C (39–77 °F)
Sample volume per measurement 120–150 mL
(approx. 20 mL per additional measuring module)

Compressed gas:
necessary quality of air if used as compressed gas supply

Class 5 from ISO 8573-1
max. particle size: 40 μm
max. pressure dew point: +7 °C (44.6 °F)
max. oil content: 25 mg/m³
Compressed gas:
necessary quality of nitrogen supply if Option O2 is installed:
Purity ≥ 99.999 %
Class 5.0
max. particle size: 40 μm
max. pressure dew point: +7 °C (44.6 °F)
Operating pressure for PBA system 6 ±0.5 bar relative (87 ±7 psi)
Dimensions (L x W x H) of the PBA system 495 x 700 x 670 mm (19.5 x 27.6 x 26.4 in)
Voltage AC 100 to 240 V, 50/60 Hz
Power consumption 190 VA (incl. external measuring modules)
Power inlet according to IEC/EN 60320-1/C14, protection class I
Fuses Glass tube fuses 5 x 20 mm; IEC60127-2; AC 250 V; T 5 AH
Environmental conditions (EN 61010) Indoor use only
Ambient temperature 15–35 °C (59–95 °F)
Air humidity 10–90 % relative humidity, non-condensing
Pollution degree 2
Overvoltage category II
Altitude max. 3000 m (9800 ft)


Technical Data of DMA™ M (all versions)


Touch screen 10.4″ TFT PCAP touch screen 640 x 480 px
Housing material Sealed housing
  • Top and side cover: Aluminum, coated
  • Back: Aluminum
  • Front: Xsample slot cover plate: Polystyrene / Butadiene
Memory 1000 measured data with/without camera pictures
Interfaces 4 x USB (2.0 full speed)
1 x Ethernet (100 Mbit)
1 x CAN Bus
1 x RS-232
1 x VGA
RS-232 printer settings Interface: RS-232 C; Baud rate: 9600; Parity: none; Stop bit: 1; Data bits: 8