Home   |   Sitemap   |   PM   |   Distributors' Corner   |   Contact
  Anton Paar GmbH  
search
   
User: login help
Key:
login
register
   
Temperature measuring instruments
A A
Detailed Information
  
Features & Benefits
Specifications
Download
  Images
   
  Millikelvin Thermometer MKT 50
 
Millikelvin Thermometer MKT 50
Enlarge image
  Millikelvin Thermometer MKT 50
 
Millikelvin Thermometer MKT 50
Enlarge image
    

Millikelvin Thermometer MKT 50
Specifications

 

 
Measuring range
Temperature:

Resistance:

-260 to +962 °C
(-418 to +1764 °F)
0 to 440 Ω

Uncertainty of measurement*
Temperature:
Resistance:

< 1 mK (Pt 100)
< 400 µΩ
*) refers to the calibration of the reference resistor without contribution of the sensor (confidence level: 95 %, number of measurement values: 50, reference temperature:23 °C)

Resolution
Temperature:
Resistance:

0.1 mK (Pt 100)
40 µΩ

Measuring sensor: Pt 100 or Pt 25.5
(DIN IEC 751 or ITS-90)

Measuring current: 0.41 mArms

Internal reference resistor: 400 Ω (Vishay)

Number of sensor inputs: 2

Sensor connection: 4 wire (Lemo 1S304)

Measuring time (for both channels):

1.44 seconds

Data outputs: RS 232 D; LAN (Ethernet, 10 Mbit)

Ambient operating temperature:

0 to 35 °C (for highest accuracy 20 to 25 °C recommended)


Temperature coefficient:

< 0.3 ppm/°C


Power supply:

Power adapter 7.5 V or 2xAA batteries


Dimensions (W x H x D):

approx. 190 x 110 x 240 mm (7.5 x 4.4 x 9.5 inches)


Weight:

approx. 2 kg (4.4 lbs)


 
    