Millikelvin Thermometer: MKT 50
In combination with two highly precise Platinium Resistance Thermometers your MKT 50 Millikelvin Thermometer measures the temperature to the highest level of accuracy. MKT 50 is by far the best-priced thermometer for applications which require a low measuring uncertainty of 0.01 °C to 0.001 °C.
Highest accuracy
- Resolution 0.0001 °C (0.1 mK)
- Measuring uncertainty over the whole measuring range 0.001 °C (= 1 mK)
- Anton Paar's patented direct current measuring method (previous models MKT 25/100)
Temperature calculation according to conventional methods
- DIN IEC 751 for industrial Platinum Resistance Thermometers (PRT)
- ITS-90 for standard thermometers (SPRT: Standard Platinum Resistance Thermometer)
- Comparison calibration (PRT) or fixed point calibration (SPRT)
- Calibration data for 30 sensors can be saved
Highest stability - easy adjustment
- AD converter does not drift
- Adjustment of the internal reference resistance by connecting a standard resistor
- Automatic new calculation of the reference resistance in less than 2 minutes
Modern interfaces plus battery and mains operation
- LAN connection (Ethernet)
- Serial interface RS-232
- Mains adapter
- Battery operation: 2 AA batteries (>10 hours)
Applications
- Highly accurate comparison calibrations of all types of thermometers in liquid baths
- Reference thermometer for all applications and branches
- Calibration of standard thermometers at temperature fixed points according to ITS-90
Industries
- Education, Research
- Electronics
- Materials Science
| Measuring range |
| Temperature |
-260 to +962 °C (-418 to +1764 °F)
|
| Resistance |
0 to 440 Ω |
Uncertainty of measurement*
|
| Temperature |
< 1 mK (Pt 100)
|
|
Resistance
|
< 400 µΩ
|
| Resolution |
| Temperature |
0.1 mK (Pt 100)
|
| Resistance |
40 µΩ
|
| Measuring sensor |
Pt 100 or Pt 25.5 (DIN IEC 751 or ITS-90) |
| Measuring current |
0.41 mArms
|
| Internal reference resistor |
400 Ω (Vishay)
|
| Number of sensor inputs |
2
|
| Sensor connection |
4 wire (Lemo 1S304) |
| Measuring time (for both channels) |
1.44 seconds |
| Data outputs |
RS 232 D (optionally USB with USB-RS232 converter); LAN (Ethernet, 10 Mbit)
|
| Ambient operating temperature |
0 to 35 °C (for highest accuracy 20 to 25 °C recommended) |
| Temperature coefficient |
< 0.3 ppm/°C |
| Power supply |
Mains adapter 7.5 V or 2 x AA batteries |
| Dimensions (L x W x H) |
approx. 240 x 190 x 110 mm (9.5 x 7.5 x 4.3 inches) |
| Weight |
approx. 2 kg (4.4 lbs) |
|
*) refers to the calibration of the reference resistor without contribution of the sensor (confidence level: 95 %, number of measurement values: 50, reference temperature: 23 °C) |